Climatic Test Chamber CTC256/Temperature Test chamber TTC256
CHF 31,080.00*
Content:
1
Delivery according to order confirmation
Product number:
9868010
Manufacturer:
Memmert GmbH + Co. KG
Weight:
421 kg
* Specifications refer to the packaging unit
Simulate the perfect atmosphere for climate and temperature tests, specifically in accordance with DIN EN 60068-2-1, 2-2 and 2-3. Provide unlimited flexibility for controlled material and function tests as well as ageing tests.
Standard equipment
Ventilation and Control:
CTC/TTC: without humidity control: from -42°C up to +190°C
CTC: with humidity control: from +10°C up to +95°C
Power supply 400 V, 3ph, N, 50 Hz
Standards for CTC additionally:
Standard equipment
Ventilation and Control:
- high-performance air fan in working chamber
- depending on operation status automatically variable fan speed with manual adjustment from 10% to 100%
- adaptive, fuzzy-supported multifunctional digital microprocessor PID-controller
- autodiagnostic system with fault indication on temperature and humidity control
- 2 x Class A, 4-wire circuit, Pt100 sensors, mutually monitoring and maintaining performance at the same temperature value
- digital 7-day-programme-timer with real time clock, precise minute setting
- integral timer for heating profiles of up to 40 ramps, each segment adjustable from 1 min. up to 999 hours
- digital display (LED) of all set parameters, such as temperature, weekdays, time, humidity and set-up values - language to be chosen in setup
- digital display of set values (resolution: 0.1°C below 99.9°C, 0.5°C above 100°C) and actual values (resolution: 0.1°C) of temperature (LED)
- long-term documenting (ring store) of all relevant data, GLP-conforming as data logger - 1024 kB
- programme stored on power failure
- parallel printer interface (incl. real-time clock with date function) for all PCL3-compatible ink-jet printers for GLP-conforming documentation
- USB interface including Memmert "Celsius" software for programming and documentation
- chip card control incl. one MEMoryCard XL with 32 kB (up to 40 ramps)
- incl. works calibration certificate for -20°C and +160°C
- high-performance ring heaters with optimised air circulation
- door heating to avoid condensate
- twin compressor
- speed adjustable condenser fan
- refrigerant R404A
- with audible and visual alarm in case of over-/undertemperature and over-/under-humidity, open door and empty water tank
- independently working, digitally adjustable electronic overtemperature controller, TWW protection class 3.3
- additional adjustable Auto-Safety-Function "ASF" for over- and undertemperature automatically following the set value at a preset tolerance range
- mechanical temperature limiter TB class 1 switching the heating off at approx.10°C above max. oven temperature
CTC/TTC: without humidity control: from -42°C up to +190°C
CTC: with humidity control: from +10°C up to +95°C
Power supply 400 V, 3ph, N, 50 Hz
Standards for CTC additionally:
- Electronically controlled active humidification and dehumidification from 10% to 98% rh with digital display of the relative humidity, display resolution 0.5%, setting accuracy 1%
- incl. works calibration certificate for +30°C at 60% rh
- Humidity supply with distilled water from 2 x 10 l tanks on integrated telescopic slide, with automatic water tank change-over
- 2 x 10L tanks acting as condensate collectors, on telescopic slide mounts
Capacity l: | 256 |
---|---|
Dual Use: | 0 |
Housing (W x D x H) mm: | 898 x 1100 x 1730 |
Internal dimensions (W x D x H) mm: | 640 x 597 x 670 |
Manufacturer Article Number: | CTC256 |
Rating W: | 7000 |
Shelf support ribs/shelves: | 6/1 |
Type: | CTC256 |
Weight kg: | 337 |